GPR-1600 Series Trace PPM 氧分析仪
产 地:美国
品 牌:AII
产品简介:
Features an advanced galvanic trace PPM oxygen sensor capable of precisely measuring oxygen concentrations ranging from 100 PPB to 1% over a 24 month operating life with no periodic maintenance required, menu driven controls, and, a stainless steel integral bypass sample system that enables the analyzer to come online at low PPM levels in minutes.
技术参数:
Accuracy: < +1% of FS range under constant conditions
Analysis: 0-10 ppm, 0-100, 0-1000, 0-1%, 0-25% FS ranges;
auto-ranging or manually lock on single range
Application: Oxygen analysis from 100 ppb to 1% in inert, helium, hydro-
gen, mixed and acid (CO2) gas streams
Approvals: CE
Area Classification: General purpose
Alarms: 2 adjustable form C relay contacts non-latching; “weak sen-
sor” indicator; power failure; system failure
Calibration: Certified gas of O2 balance N2 approximating 80% of analysis
range or one range higher than analysis range
Compensation: Barometric pressure and temperature; optional temperature
controlled heated sample system
Connections: 1/4" compression tube fittings
Controls: Water resistant keypad; menu driven range selection, cali-
bration, alarm and system functions
Data Acquisition: Selectable data point intervals
Display:
Graphical LCD 5 x 2.75; resolution .01 ppm; displays real
time ambient temperature and pressure
Enclosure: Painted aluminum 7.5” x 10.8” x 12.25” panel mount
Flow Sensitivity: None between 1-5 SCFH, 2 SCFH recommended
Linearity: > .995 over all ranges
Pressure:
Inlet - regulate to 5-30 psig, max 100 psig;
vent - atmospheric not to exceed –14” water column
Power: Universal; specify 110 or 220 VAC with optional heater
Range ID: 1-5V output with 1V representing the lowest range
Recovery Time: 60 seconds in air to < 10 ppm in < 1 hr on N2 purge
Response Time: 90% of final FS reading < 10 seconds
Sample System: Wetted parts: stainless steel consisting of flow control and
sample/bypass valves; flow indicator
Sensitivity: < 0.5% of FS range
Sensor Model: GPR-12-333 - requires no maintenance
Sensor Life: 24 months at 25ºC and 1 atm; average O2 < 1000 ppm
Signal Output: 4-20mA isolated and 0-1V
Temp. Range: 5º to 45ºC (GPR sensor); -20º to 45ºC (XLT sensor)
Warranty: 12 months analyzer; 12 months sensor
先进的传感技术:
Accuracy < +1% FS Range
Sensitivity < 0.5% FS Range
Fast Recovery to < 10 ppm
24 Month Expected Life
No Maintenance
Compatible in 0-100% CO2
Auto Zero, Span Calibration
5 Standard Ranges
Auto Ranging & Manual Selecti
Remote Communication Link
Certified ISO 9001 QA System