Enclosure |
Class 1, Div1 Grps C & D |
Operating Principle |
Dual-beam, non-dispersive infrared (NDIR) |
Detector |
Sapphire window with glass frit seal |
Sensing Cell Proof Pressure (sample cell only) |
2 PSI (13.8 KPa) gauge pressure |
Measurement Range |
0–20% CO2 (other ranges available upon request) |
Repeatability |
± 0.1 % CO2 |
Lower Detectable Limit |
0.1 % CO2 |
Drift |
Less than 2.5% of reading/year |
Accuracy |
± 3% of reading or 0.1% CO2 at cal temperature ± 5% of reading or 0.1% CO2 over full operating temperature range |
Warmup time |
Less than 5 minutes |
Operating Temperature |
10 - 50˚C |
Operating Humidity |
Sensing Cell: 0–100% RH (non condensing)
Electronics: 0–90% RH (non condensing) |
Storage Temperature |
-20–60˚C. |
Storage Humidity |
0–90% RH (non condensing) |
Analog Output |
Isolated 4–20 mA |
Digital Interface |
RS-485 & RS-232 |
Operating Power |
24 VDC at 5 Watts standard, 12 VDC optional |
Source Life |
10 years minimum |
Alarm Outputs |
2 solid state drivers 2 Amp, 30 VDC maximum |
Standard Sample System |
Inlet filter (1,500 psig max) and SS single stage regulator (3,000 psig max), flow meter (100 psig max) mounted |